5 matches found
CVE-2022-38069
CVE-2022-38069 concerns Contec Health CMS8000 devices. The root cause is multiple globally default credentials present across all CMS8000 units. If exposed, a threat actor with momentary physical access can gain privileged access to any device, enabling extraction of sensitive patient information...
CVE-2022-38100
The CVE-2022-38100 vulnerability affects Contec Health CMS8000 (and CME8000 devices) where parsing malformed network data causes a crash requiring a physical reboot. A threat actor with network access can remotely issue a specially formatted UDP request, and a UDP broadcast could trigger a mass d...
CVE-2022-3027
CVE-2022-3027 affects Contec Health CMS8000 CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor. The issue is an improper control/sanitization of the SSID name for a new Wi‑Fi access point, allowing an attacker to craft an SSID with malicious/non‑standard characters. When the device attempts to co...
CVE-2022-36385
CVE-2022-36385 affects Contec Health CMS8000 family (CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor). The vulnerability arises from improper access controls that permit a threat actor with momentary physical access to insert a USB drive and perform a malicious firmware update, causing permane...
CVE-2022-38453
The CVE-2022-38453 issue affects the CMS8000 device, where multiple binary application files are compiled with 'not stripped' and 'debug_info' settings. This weakens reverse-engineering resistance and could enable an attacker to identify additional vulnerabilities in the affected software stack. ...